Volume 43 Issue 11
Nov.  2022
Turn off MathJax
Article Contents
Dongmei YANG, Chunyan XIA, Tingjun LUO, Yixiao WANG, Zhen JIN. Establishment and Application Evaluation of Automatic Verification System for Coagulation Test Reports[J]. Journal of Kunming Medical University, 2022, 43(11): 58-62. doi: 10.12259/j.issn.2095-610X.S20221107
Citation: Dongmei YANG, Chunyan XIA, Tingjun LUO, Yixiao WANG, Zhen JIN. Establishment and Application Evaluation of Automatic Verification System for Coagulation Test Reports[J]. Journal of Kunming Medical University, 2022, 43(11): 58-62. doi: 10.12259/j.issn.2095-610X.S20221107

Establishment and Application Evaluation of Automatic Verification System for Coagulation Test Reports

doi: 10.12259/j.issn.2095-610X.S20221107
  • Received Date: 2022-06-24
    Available Online: 2022-10-28
  • Publish Date: 2022-11-14
  •   Objective  To establish an automatic review system for blood coagulation test report and evaluate its application effect.   Methods  The rules for the examination of coagulation test reports were written into Lis Procedure. A total of 6281 reports that were automatically audited during the commissioning period were selected for manual verification, to evaluate the feasibility of the rules.   Results  The passing rate of automatic verification was 71.34%, False positive rate (4.49%), False negative rate (0.032%), Positive coincidence rate (24.14%), Negative coincidence rate (71.34%), Consistent rate (95.48%) and correct passing rate (99.96%).   Conclusion  There are 23 intercepting rules, the Delta Check rule of APTT needs to be improved, and the rest of 22 intercepting rules’ correct passing rate is 100%.
  • loading
  • [1]
    Autoverification of Clinical Laboratory Test Results;Approved Cuide-line[S]. AUTO-10A, CLSI, 2006.
    [2]
    Randell E W,Yenice S,Khine Wamono A A,et al. Autoverification of test results in the core clinical laboratory[J]. Clin Biochem,2019,73:11-25.
    [3]
    冯强. 临床生化检验结果的自动审核[J]. 国际医学检验杂志,2015,36(4):547-548.
    [4]
    杨勇,贾延伟,曹伟,等. 生化检验结果自动审核规则的制定及其应用[J]. 临床检验杂志,2016,34(2):152-155.
    [5]
    蒋文海,吴炯,阴忆青,等. 临床化学实验室自动审核系统搭建策略及优化[J]. 中华临床实验室管理电子杂志,2015,3(2):96-100. doi: 10.3877/cma.j.issn.2095-5820.2015.02.008
    [6]
    雷和月,王红旗,郭远瑜,等. 免疫学检验结果自动审核程序的建立及其应用[J]. 临床检验杂志,2015,33(5):388-390.
    [7]
    王文洋,冯强,王超群,等. 血常规自动审核模式的改进[J]. 中华临床实验室管理电子杂志,2017,5(4):236-239. doi: 10.3877/cma.j.issn.2095-5820.2017.04.009
    [8]
    尚红, 王毓三, 申子瑜. 全国临床检验操作规程[M]. 第4版. 北京: 人民卫生出版社, 2015: 96-117.
    [9]
    王剑飚. 血细胞分析系统自动审核规则建立与确认[J]. 中国医疗器械信息,2016,13(7):7.
    [10]
    曲琳琳,吴俊. 凝血检验结果自动审核规则建立与验证的多中心研究[J]. 中华检验医学杂志,2020,43(8):806-809.
    [11]
    李玲,杜佳,赵耿生,等. 凝血试验自动审核方案的建立与临床应用[J]. 临床检验杂志,2017,39(7):536.
    [12]
    温冬梅. 临床实验室生化免疫自动审核系统的建立及应用[J]. 中华检验医学杂志,2018,41(2):141-148. doi: 10.3760/cma.j.issn.1009-9158.2018.02.013
    [13]
    夏良裕,程歆琦,刘茜,等. 临床实验室生化免疫项目自动审核程序的建立与应用[J]. 中华医学杂志,2017,97(8):616-621. doi: 10.3760/cma.j.issn.0376-2491.2017.08.012
    [14]
    续薇,郝晓柯,崔巍,等. 血液分析自动审核规则建立与验证的多中心研究[J]. 中华检验医学杂志,2018,41(8):601-607. doi: 10.3760/cma.j.issn.1009-9158.2018.08.009
    [15]
    中华人民共和国国家卫生健康委员会. 临床实验室定量检验结果的自动审核: WS/T616-2018[S]. 北京: 中国标准出版社, 2018.
    [16]
    朱晶,潘柏申. 临床检验结果自动审核应用进展[J]. 临床检验杂志,2018,36(12):889. doi: 10.13602/j.cnki.jcls.2018.12.02
  • Relative Articles

    [1] Qingmei YU, Tao SHI, Hongxue GUAN, Yujun MAO, Zhaoyi LI, Jiao YANG. Prevalence and Influence Factors of Renal Function Impairment in Patients Following Radical Cystectomy:A Systematic Review and Meta-analysis. Journal of Kunming Medical University, 2024, 45(9): 83-90.  doi: 10.12259/j.issn.2095-610X.S20240913
    [2] Xinyue ZHANG, Guangmin ZHENG, Rui YANG, Zhongyan ZHANG, Minxuan SHEN, Jing CHEN. Development and Application of Clinical Case Mind Map Library Based on System Organs. Journal of Kunming Medical University, 2024, 45(8): 166-173.  doi: 10.12259/j.issn.2095-610X.S20240824
    [3] Yuqin LIU, Mingying YANG, Liufang WANG, Siya XIA, Danna LI, Xijuan ZHAO, Ya WANG. Development of a Patient-reported Outcome Scale For Bladder Cancer and Test of Reliability and Validity. Journal of Kunming Medical University, 2024, 45(7): 183-190.  doi: 10.12259/j.issn.2095-610X.S20240727
    [4] Yu LUO, Shinan ZHANG, Ni ZHOU, Changhai LYU. Application of Different Sealing Materials and Adhedive Systems in Pit and Fissure Sealing of Deciduous Molars. Journal of Kunming Medical University, 2023, 44(3): 74-80.  doi: 10.12259/j.issn.2095-610X.S20230314
    [5] Yueyue TANG, Hanning YANG, Youliang JIN, Shumin LI, Yongping LU. Clinical Investigation on Foure-dimensional Left Atrial Automatic Quantitative Technology in Evaluating Left Atrial Volume and Function. Journal of Kunming Medical University, 2023, 44(7): 57-63.  doi: 10.12259/j.issn.2095-610X.S20230713
    [6] Shunding TANG, Chonghua WAN, Ying SONG, Li DENG, Han WANG, Changwei SU, Xiaoqing ZHANG. Evaluation and Application of the Patients Reported Outcomes Measurement Scale for Hypertension. Journal of Kunming Medical University, 2022, 43(3): 60-66.  doi: 10.12259/j.issn.2095-610X.S20220312
    [7] Yun-song XUE, Li-li LI, Qian FENG, Yi-zheng LI, Lin-hua ZHOU, Jin-yi GU, Jian-mei LIU. Evaluation of Analysis Performance of Automatic Biochemical Analyzer in Detecting Serum Lipoprotein A. Journal of Kunming Medical University, 2021, 42(3): 49-53.  doi: 10.12259/j.issn.2095-610X.S20210312
    [8] Mei Song , Jiang Wen , Bai Xiang Feng , Tao Jie , Wang Wen Jie , Li Jia Yu . . Journal of Kunming Medical University, 2020, 41(06): 85-88.
    [9] Li Liu Zheng , Wang Jun Feng , Luo Ming Ju , Zhang Hong Bo , Li Bo , Bao Bing Bing , Li Yi , Xu Lei Sheng . . Journal of Kunming Medical University, 2018, 39(11): 87-92.
    [10] Li Gui Mei , Chen Wei Wen , Xu Ying Bo , Yin Xiao E . Comparative Analysis of Test Results between POCT Blood Glucose Meter and Automatic Biochemical Analyzer. Journal of Kunming Medical University, 2017, 38(08): 106-109.
    [11] Yao Jin Yong , Hu Li Ping , Nie Ai Ting , Huang Ren Wu , Qiu Jian Bo , Zhang Xiu Feng , Rao 旼, Nie Sheng Jie . Establishment and Application An Automatic Typing Method for MAOA μ-VNTR Polymorphism Test using Fluorescence Labeling Techniques. Journal of Kunming Medical University, 2017, 38(06): 52-55.
    [12] Wang Sheng Hu . Comparison of the Test Results between Automatic Enzyme-linked Immunity Analyzer and Manual Enzyme-linked Immunoassay. Journal of Kunming Medical University,
    [13] Wen Lin . . Journal of Kunming Medical University,
    [14] Feng Qing . . Journal of Kunming Medical University,
    [15] Liu Hai Bo . . Journal of Kunming Medical University,
    [16] Chen Jian . . Journal of Kunming Medical University,
    [17] Yang Shun Qiu . . Journal of Kunming Medical University,
    [18] Deng De Yao . . Journal of Kunming Medical University,
    [19] Diao Chang . . Journal of Kunming Medical University,
    [20] Cheng Si Gen . . Journal of Kunming Medical University,
  • 加载中

Catalog

    通讯作者: 陈斌, bchen63@163.com
    • 1. 

      沈阳化工大学材料科学与工程学院 沈阳 110142

    1. 本站搜索
    2. 百度学术搜索
    3. 万方数据库搜索
    4. CNKI搜索

    Figures(2)  / Tables(3)

    Article Metrics

    Article views (2652) PDF downloads(120) Cited by()
    Proportional views
    Related

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return
    Return