Yi ZHONG, Zheng-yang ZHOU. Application of Coronary Computed Tomography Angiography in Overweight Patients: Comparison of Automated Tube Voltage Selection Versus 100 kVp Tube Voltage[J]. Journal of Kunming Medical University, 2021, 42(3): 35-40. doi: 10.12259/j.issn.2095-610X.S20210310
Citation: Yi ZHONG, Zheng-yang ZHOU. Application of Coronary Computed Tomography Angiography in Overweight Patients: Comparison of Automated Tube Voltage Selection Versus 100 kVp Tube Voltage[J]. Journal of Kunming Medical University, 2021, 42(3): 35-40. doi: 10.12259/j.issn.2095-610X.S20210310

Application of Coronary Computed Tomography Angiography in Overweight Patients: Comparison of Automated Tube Voltage Selection Versus 100 kVp Tube Voltage

doi: 10.12259/j.issn.2095-610X.S20210310
  • Received Date: 2020-12-11
  • Publish Date: 2021-03-25
  •   Objective   To compare the image quality and radiation dose of automated tube voltage selection (APSCM) and 100 kVp coronary computed tomography angiography (CCTA) in overweight patients.   Methods   A total of 115 patients with a body mass index (BMI) higher than 25 kg/m2 and less than 30 kg/m2 and body weight lower than 90 kg underwent high-pitch prospectively ECG-triggered CCTA on a second-generation dual-source CT scanner. All patients were classified into two groups: Group A, 75 patients were examined with APSCM; Group B, 40 patients with a fixed tube voltage of 100 kVp. The attenuation and image noise were measured and signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) were calculated. Image quality was evaluated and radiation dose was recorded.   Results   In Group A, 80 kVp, 100 kVp and 120 kVp were automatically selected in 1 (1.3%), 51 (68.0%) and 23 (30.7%) patients, respectively. The mean CT attenuations and image noise in APSCM group were significantly lower than those in Group B (all P < 0.001). Group A showed a higher mean CNR but a lower mean SNR (all P < 0.05) than Group B. The subjective image quality scores have no significant difference between the two groups (all P < 0.05). The radiation dose in Group B was 22.2% lower than in Group A ( P = 0.004).   Conclusion   In patients with BMI higher than 25 kg/m2 and less than 30 kg/m2, 100 kVp CCTA allows lower radiation dose without compromising image quality when compared with CCTA using APSCM.
  • loading
  • [1]
    BamiK,Premaratne M,Lamba J,et al. Appropriate use criteria for cardiac computed tomography:impact on diagnostic utility[J]. Journal of Computer Assisted Tomography,2017,41(5):1.
    [2]
    Richards C E,Dorman S,John P,et al. Low-radiation and high image quality coronary computed tomography angiography in "real-world" unselected patients[J]. World J Radiol,2018,10(10):135-142. doi: 10.4329/wjr.v10.i10.135
    [3]
    Bosch de Basea M,Pearce M S,Kesminiene A,et al. EPI-CT:design,challenges and epidemiological methods of an international study on cancer risk after paediatric and young adult CT[J]. Journal of Radiological Protection,2015,35(3):611-628. doi: 10.1088/0952-4746/35/3/611
    [4]
    Mccollough C H. Computed tomography technology-and dose-in the 21st century[J]. Health Physics,2019,116(2):157-162. doi: 10.1097/HP.0000000000000997
    [5]
    Abbara S,Blanke P,Maroules C D,et al. SCCT guidelines for the performance and acquisition of coronary computed tomographic angiography:A report of the society of Cardiovascular Computed Tomography Guidelines Committee:Endorsed by the North American Society for Cardiovascular Imaging (NASCI)[J]. J Cardiovasc Comput Tomogr,2016,10(6):435-449.
    [6]
    Hecht H S,Cronin P,Blaha M J,et al. 2016 SCCT/STR guidelines for coronary artery calcium scoring of noncontrast noncardiac chest CT scans:A report of the Society of Cardiovascular Computed Tomography and Society of Thoracic Radiology[J]. J Cardiovasc Comput Tomogr,2017,11(1):74-84.
    [7]
    Den Harder A M,Willemink M J,de Jong P A,et al. New horizons in cardiac CT[J]. Clin Radiol,2016,71(8):758-767. doi: 10.1016/j.crad.2016.01.022
    [8]
    Lee H S,Suh Y J,Han K,et al. Effectiveness of automatic tube potential selection with tube current modulation in coronary CT angiography for obese patients:Comparison with a body mass index-based protocol using the propensity score matching method[J]. PLoS One,2018,3(1):e0190584.
    [9]
    Zhao P,Hou Y,Liu Q,et al. Radiation dose reduction in cardiovascular CT angiography with iterative reconstruction (AIDR 3D) in a swine model:A model of paediatric cardiac imaging[J]. Clin Radiol,2016,71(7):716.
    [10]
    Örgel A,Bier G,Hennersdorf F,et al. Image quality of CT angiography of supra-aortic arteries:comparison between advanced modelled iterative reconstruction (ADMIRE),sinogram affirmed iterative reconstruction (SAFIRE) and filtered back projection (FBP) in one patients' group[J]. Clin Neuroradiol,2020,30(1):101-107. doi: 10.1007/s00062-018-0740-y
    [11]
    Andreini D,Pontone G,Mushtaq S,et al. Image quality and radiation dose of coronary CT angiography performed with whole-heart coverage CT scanner with intra-cycle motion correction algorithm in patients with atrial fibrillation[J]. Eur Radiol,2018,28(4):1383-1392. doi: 10.1007/s00330-017-5131-2
    [12]
    Mousavi Gazafroudi S S,Tavakkoli M B,Moradi M,et al. Coronary CT angiography by modifying tube voltage and contrast medium concentration:Evaluation of image quality and radiation dose[J]. Echocardiography,2019,36(7):1391-1396. doi: 10.1111/echo.14410
    [13]
    Atwi N E,Smith D L,Flores C D,et al. Dual-energy CT in the obese:a preliminary retrospective review to evaluate quality and feasibility of the single-source dual-detector implementation[J]. AbdomRadiol (NY),2019,44(2):783-789.
    [14]
    Symons R,Choi Y,Cork T E,et al. Optimized energy of spectral coronary CT angiography for coronary plaque detection and quantification[J]. J Cardiovasc Comput Tomogr,2018,12(2):108-114.
    [15]
    Austen W G,Edwards J E,Frye R L,et al. A reporting system on patients evaluated for coronary artery disease:report of the ad hoc committee for grading of coronary artery disease,council on cardiovascular surgery,American Heart Association[J]. Circulation,1975,51(Suppl4):5-40.
    [16]
    Lee S H,Cho J H,Kim H E,et al. Effect of coronary artery calcification score by lifestyle and correlation with coronary artery stenosis by multidetector computed tomography[J]. J Comput Assist Tomogr,2017,41(2):236-241. doi: 10.1097/RCT.0000000000000497
    [17]
    Dorbala S,Blankstein R,Skali H,et al. Approaches to reducing radiation dose from radionuclide myocardial perfusion imaging[J]. J Nucl Med,2015,56(4):592-599. doi: 10.2967/jnumed.112.115097
    [18]
    Kawashima H,Ichikawa K,Hanaoka S,et al. Relationship between size-specific dose estimates and image quality in computed tomography depending on patient size[J]. J Appl Clin Med Phys,2018,19(4):246-251. doi: 10.1002/acm2.12340
    [19]
    Lee J W,Kim C W,Lee H C,et al. High-definition computed tomography for coronary artery stents:image quality and radiation doses for low voltage (100 kVp) and standard voltage (120 kVp) ECG-triggered scanning[J]. Int J Cardiovasc Imaging,2015,31(Suppl1):39-49.
    [20]
    Lee H S,Suh Y J,Han K,et al. Effectiveness of automatic tube potential selection with tube current modulation in coronary CT angiography for obese patients:Comparison with a body mass index-based protocol using the propensity score matching method[J]. PLoS One,2018,13(1):e0190584. doi: 10.1371/journal.pone.0190584
    [21]
    Li M,Feng S,Wu N,et al. Scout-based automated tube potential selection technique (kV assist) in enhanced chest computed tomography:Effects on radiation exposure and image quality[J]. J Comput Assist Tomogr,2017,41(3):442-445. doi: 10.1097/RCT.0000000000000523
  • Relative Articles

    [1] Hui WANG, Wenjun ZENG, Haiping ZHANG, Ruiwei GUO. Effects of Store-operated Calcium Channel Molecule Orai3 on Rat Coronary Artery Smooth Muscle Cells Proliferation. Journal of Kunming Medical University, 2023, 44(4): 16-22.  doi: 10.12259/j.issn.2095-610X.S20230423
    [2] Juan ZHAO, Teng-can TU, Mei-jun YANG, Xiao-yong ZHANG, Wei PAN. Assessment of Early Coronary Artery Disease in Patients with Acute Coronary Syndrome. Journal of Kunming Medical University, 2021, 42(12): 47-50.  doi: 10.12259/j.issn.2095-610X.S20211219
    [3] Yang Bin , Du Wei , Duan Wen Shuai . . Journal of Kunming Medical University, 2020, 41(09): 114-119.
    [4] Ren Chao Feng , Yang Li Fen , Liu Fen , Zhao Ya Lin , Li Mei Hua . . Journal of Kunming Medical University, 2019, 40(10): 146-149.
    [5] Wang Shu Ying , Liu Guang Yu , Zhang Ling , Li Zhi Chun , Lin Yi Qiu , Sheng Bo . . Journal of Kunming Medical University, 2018, 39(09): 115-119.
    [6] Yuan Bin . Roles of Percutaneous Coronary Intervention in Improving Left Ventricular Function, Reducing Incidence of Cardiovascular Events of Patients With Ischemic Cardiomyopathy. Journal of Kunming Medical University,
    [7] Jia Xiao Xia . . Journal of Kunming Medical University,
    [8] Wu Yan . . Journal of Kunming Medical University,
    [9] Ma Min . . Journal of Kunming Medical University,
    [10] Feng Zhen Xia . . Journal of Kunming Medical University,
    [11] Wu Jian . . Journal of Kunming Medical University,
    [12] Chen Ting Jie . . Journal of Kunming Medical University,
    [13] He Yan . . Journal of Kunming Medical University,
    [14] Yang Li . . Journal of Kunming Medical University,
    [15] Wu Yan . . Journal of Kunming Medical University,
    [16] Peng Min An . . Journal of Kunming Medical University,
    [17] Li Fei . . Journal of Kunming Medical University,
    [18] Min Jie Qing . . Journal of Kunming Medical University,
    [19] Yang Fang . . Journal of Kunming Medical University,
    [20] . . Journal of Kunming Medical University,